Dosimetric Analysis
The PDD data was measured with the CC13 pin-point
ionization chamber detector for the different field sizes as
shown in figure
1. The larger field sizes have higher PDD
values for our dedicated system, the variation in PDD is
limited to 3% between the 3 cm x 3 cm and 40 cm x 40
cm open field sizes at depth of 10 cm.
Fig 1: Percentage depth dose of respective open field sizes using CC13 pin-point
ion chamber
The percentage depth dose for wedged field sizes was
measured and the variation found to be 4.2% between the
3 cm x 3 cm and 30 cm x 40 cm wedged field sizes at depth
of 10 cm as shown in figure 2.
Fig 2: Percentage depth dose of respective wedged field sizes using CC13 pin-point
ion chamber
The OAR were computed from the beam profiles for
10 cm x 10 cm, 20 cm x 20 cm, and 30 cm x 30 cm field
sizes at 10 and 20 cm depths in the x (crossline) and y
(inline) directions and were found within ±0.2 mm. The
OAR values entered in the treatment planning system were the average of the x and y profiles. The x and y profiles
were as shown in figure 3a and figure 3b. The wedged
profiles were measured at 1.5 cm, 5 cm, 10 cm, and 20 cm
depths as shown in Figure 3c.
Fig 3a: Crossline dose profile
Fig 3b: Inline dose profile
Fig 3c: The wedged beam profiles measured at 1.5 cm, 10 cm, and 20 cm depths
The relative output factors (OF) for different field sizes
was as shown in figure 4 and the scatter factor Sc(r) for
open field sizes was as shown in figure 5.
Fig 4: The relative output factors for various field sizes
Figure 5: Scatter factors for open field sizes
Quality-assurance of dose calculation by
measurements
Point dose calculations
The absolute point dose of the treatment plan was
measured by using Head and Neck Cube supplied by
Scanditronix Wellhofer. The dimension of this cube was
18 cm x 18 cm x 18 cm and had the facility to insert the
ion chamber. The prostate plan was executed on this
phantom and the point dose at the isocenter was calculated
(TPS value). The same plan was executed on machine with
same phantom by inserting 0.65 cc active volume Farmer
type ion chamber (FC65-G). The dose at isocenter was
measured. The calculated and the measured doses were
compared and the variation was noted in table 1.
Table 1: Comparison of isocenter doses between the calculated the measured doses for 5 IMRT plans
The deviations between the calculated dose, Dcalc, and
the measured dose, Dmeas at a specific depth was given by
(Dcalc – Dmeas) x 100% / Dmeas. In those cases where the
points were outside the penumbra or under a block, the
results of the comparison were expressed relatively to the
dose measured at the same depth, but on the central axis
of the open beam, Dmeas,cax according to (Dcalc –
Dmeas)x100% / Dmeas,cax [7,8].
Isodose distribution comparison
Film measurement was performed for a prostate IMRT
plan. Same prostate plan with seven non-coplanar beams
were used. The Torso Phantom (Scanditronix Wellhofer’s)
was used to compare the calculated and the measured dose
distributions. The film (Kodak EDR) was stacked in between
the two slabs. Isodose distributions were measured using
Kodak EDR film. The exposed film was scanned with
Vidar Scanner and the distributions were seen through
Omnipro IMRT Software. The comparisons between the
calculated and the measured dose distributions were made
within Omnipro IMRT Software system [9]. The results
for the prostate IMRT plan were as shown in figures 6A-
6D.
Fig 6A: Planned image generated from TPS for a prostate case
Fig 6B: Actual exposed image with Torso Phantom for a prostate case
Fig 6C: Profiles comparison for prostate case
Fig 6D: Comparison of isodoses for prostate case
From the figures shown in Figure 6B-6C, it is seen that the
agreement between the calculated and the measured dose
is within 2% in the high-dose region. For clinical patient
specific QA we specify 3% dose difference and 3 mm
distance acceptance scaling criteria [10].
The bulk leaf transmission and back up jaw transmission
were 0.02% and 0.105%.
Treatment-planning system configuration parameters
must be measured precisely. All the dosimetric tests have
proved very useful and detected only minor deviations. The
TPS calculated dose and the measured absolute dose should
not deviate more than 3% to ensure safe treatment. It is
necessary to maintain strict criteria to compare measured
and calculated values. The patient pretreatment QA requires
significant machine time for measurements and must be completed before treatment starts. In order to reduce patient
pretreatment QA time, an independent monitor unit
calculation program can be evaluated. The introduction of
new analyzing tools, such as DTA (distance to agreement)
or g (gamma factor), correlation coefficient can be useful
to better quantify the comparison between measured versus
calculated dose distributions [11,12]. The implementation
of IMRT must not be underestimated. Every institution
should adopt a QA protocol. The QA protocol presented
here has proved adequate, and with it we have had no patient complications attributed to IMRT delivery.
ACKNOWLEDGEMENT
The authors would like to thank Vikash Pathak of Elekta Instrument (India) for his valuable suggestions and technical support to perform quantitative comparative analysis.